Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Omega 872A: (1 year) CAL VER /5500 DATE: 08-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.4 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 14 NUMBER OF LINES: 70 CONFIGURATION: Fluke 5500A ============================================================================= # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F V 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: 30 min. - REQUIRED 1.004 DISP Temperature: 22 - 24 degC 1.004 DISP Humidity: <80% 1.005 HEAD INITIAL SETUP 1.006 DISP Turn UUT on, if Battery indicator illuminates, 1.006 DISP replace battery before proceeding. 1.007 DISP {TC1 INPUT} 1.008 DISP Set TC to TC1. 1.009 DISP Connect the 5500A TC terminals to the TC1 Input using 1.009 DISP J Type thermocouple wire and connectors. 1.010 DISP Set the UUT to the 1400degF Range. 1.011 HEAD TC1 INPUT: {1400degF Range} 1.012 5500 -35degF 0.25% 1.8U _J TC 2W 2.001 5500 32degF 0.25% 1.8U _J TC 2W 3.001 5500 1390degF 0.25% 1.8U _J TC 2W 4.001 HEAD TC1 INPUT: {200degF Range} 4.002 DISP Set the UUT to the 200degF Range. 4.003 5500 -35.0degF 0.25% 1.8U _J TC 2W 5.001 5500 32.0degF 0.25% 1.8U _J TC 2W 6.001 5500 195.0degF 0.25% 1.8U _J TC 2W 7.001 HEAD TC1 INPUT: {760degC Range} 7.002 DISP Set the UUT to the 760degC Range. 7.003 5500 -35degC 0.25% 1U _J TC 2W 8.001 5500 0degC 0.25% 1U _J TC 2W 9.001 5500 750degC 0.25% 1U _J TC 2W 10.001 HEAD TC1 INPUT: {200degC Range} 10.002 DISP Set the UUT to the 200degC Range. 10.003 5500 -35.0degC 0.25% 1U _J TC 2W 11.001 5500 0.0degC 0.25% 1U _J TC 2W 12.001 5500 195.0degC 0.25% 1U _J TC 2W 13.001 HEAD {TC2 INPUT} 13.002 DISP Set TC to TC2. 13.003 DISP Connect the 5500A TC terminatls to the TC2 Input using 13.003 DISP J Type thermocouple wire and connectors. 13.004 HEAD TC1 INPUT: {1400degF Range} 13.005 DISP Set UUT to the 1400degF Range. 13.006 5500 32degF 2U _J TC 2W 14.001 END